Quantification of gentamicin in Mueller-Hinton agar by high-performance liquid chromatography

J Chromatogr B Biomed Sci Appl. 2001 Mar 25;753(1):151-6. doi: 10.1016/s0378-4347(00)00460-6.

Abstract

The aim of this study was to optimise a method for gentamicin determination in an agar matrix and to investigate if and how agar composition can affect the gentamicin diffusion kinetics during the agar diffusion tests for antibiotics sensitivity. Gentamicin was separated by RP-HPLC and detected at 365 nm after pre-column derivatization with 1-fluoro-2,4-dinitrobenzene. Recovery (> or = 79%), linearity (r2 > or = 0.997) and sensitivity (1 microg/ml) were assessed using four different agar matrices. The kinetics of gentamicin diffusion tested on BioMerieux and DID manufacturers' products showed in uninoculated agar plates significant differences that were even more pronounced in the presence of Pseudomonas aeruginosa metabolism.

MeSH terms

  • Agar
  • Chromatography, High Pressure Liquid / methods*
  • Diffusion
  • Gentamicins / analysis*
  • Gentamicins / metabolism
  • Kinetics
  • Pseudomonas aeruginosa / metabolism
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Gentamicins
  • Agar