Comparison of image quality between conventional and low-dose nonenhanced head CT

AJNR Am J Neuroradiol. 2004 Apr;25(4):533-8.

Abstract

Background and purpose: Increasing use of CT for evaluating neurologic disease may expose patients to considerable levels of ionizing radiation. We compared the image quality of low-mAs head CT scans with that of conventional nonenhanced scans.

Methods: Conventional head CT scans were obtained in 20 patients (all >65 years with history of non-CNS malignancy) by using a multidetector technique: 170 mA and 1-second scanning time (ie, 170 mAs), 140 kVp, table speed of 7.5 mm per rotation, pitch of 0.75, section thickness of 5 mm, and field of view of 25 mm. A limited volume helical data acquisition covering four 5-mm-thick images was obtained by using 90 mAs but otherwise the same parameters. Three neuroradiologists visually rated the resulting images for quality in a blinded comparison. Representative 1- to 4-mm(2) regions of interest were chosen in gray matter and white matter locations. Conspicuity and the contrast-to-noise ratio were analyzed. Statistical comparisons were done by using the Student t test.

Results: Mean gray matter conspicuity was not significantly different between the 170- and 90-mAs groups (0.39 +/- 0.19 vs 0.41 +/- 0.03, P =.32). Mean gray matter contrast-to-noise ratio was approximately 22% higher with 170 mAs than with 90 mAs (1.77 +/- 0.52 vs 1.39 +/- 0.38, P =.005). All 90-mAs images were rated as having slightly greater image noise than the 170-mAs scans but with sufficient perceived resolution.

Conclusion: Although 90-mAs head CT images were moderately noisier than 170-mAs images, they were rated as having acceptable diagnostic quality.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Artifacts
  • Brain / diagnostic imaging*
  • Brain / radiation effects
  • Female
  • Head and Neck Neoplasms / diagnostic imaging*
  • Humans
  • Male
  • Quality Control
  • Radiation Dosage
  • Radiographic Image Enhancement
  • Radiographic Magnification / standards
  • Radiometry / standards*
  • Sensitivity and Specificity
  • Tomography, Spiral Computed / standards*
  • Tomography, X-Ray Computed / standards*