Background: Dysfunction of haemodialysis catheters is most commonly due to a narrowing of the catheter lumen and/or formation of a fibrin sheath around the catheter tip. Reported methods for restoring patency of the catheter lumen include passage of a J-tipped guide wire, passage of a biopsy brush through the catheter, or infusion of a thrombolytic agent into the catheter. While these methods are often effective, they suffer from several limitations. We present a minimally invasive technique to remove thrombi and debris from within the lumen of a partially thrombosed haemodialysis catheter while simultaneously stripping the fibrous sheath.
Methods: A 0.089 cm nitinol wire is bent to create a loop, which is then inserted via the catheters. Upon exiting the lumen of the catheters, the nitinol wire forces a snare open, which disrupts the fibrin sheath and catches intraluminal thrombi and debris. The technique requires no anaesthesia or recovery time.
Results: Initial clinical success in our series was achieved in all patients (7/7) as evidenced by restoration of target flow rates on subsequent haemodialysis. None of the patients experienced any complications as a result of the procedure. The catheter 2-, 4-, and 6-week primary success rates were 100% (8/8), 100% (8/8), and 100% (8/8) respectively with a mean duration of 17.1 weeks (range 8-40 weeks).
Conclusions: The internal snare technique is an effective, inexpensive and minimally invasive approach to restoring patency to failed central venous access catheters.