Quantitative spatial distribution of sirolimus and polymers in drug-eluting stents using confocal Raman microscopy

J Biomed Mater Res A. 2008 Apr;85(1):258-70. doi: 10.1002/jbm.a.31535.

Abstract

Raman spectroscopy was used to differentiate each component found in the CYPHER Sirolimus-eluting Coronary Stent. The unique spectral features identified for each component were then used to develop three separate calibration curves to describe the solid phase distribution found on drug-polymer coated stents. The calibration curves were obtained by analyzing confocal Raman spectral depth profiles from a set of 16 unique formulations of drug-polymer coatings sprayed onto stents and planar substrates. The sirolimus model was linear from 0 to 100 wt % of drug. The individual polymer calibration curves for poly(ethylene-co-vinyl acetate) [PEVA] and poly(n-butyl methacrylate) [PBMA] were also linear from 0 to 100 wt %. The calibration curves were tested on three independent drug-polymer coated stents. The sirolimus calibration predicted the drug content within 1 wt % of the laboratory assay value. The polymer calibrations predicted the content within 7 wt % of the formulation solution content. Attenuated total reflectance Fourier transform infrared (ATR-FTIR) spectra from five formulations confirmed a linear response to changes in sirolimus and polymer content.

Publication types

  • Comparative Study

MeSH terms

  • Anti-Bacterial Agents / administration & dosage*
  • Anti-Bacterial Agents / chemistry
  • Drug-Eluting Stents*
  • Microscopy, Confocal
  • Pharmaceutical Preparations / administration & dosage*
  • Pharmaceutical Preparations / chemistry
  • Polymers* / chemistry
  • Sirolimus / administration & dosage*
  • Sirolimus / chemistry
  • Spectrum Analysis, Raman*

Substances

  • Anti-Bacterial Agents
  • Pharmaceutical Preparations
  • Polymers
  • Sirolimus