Strategy to eliminate inappropriate shocks secondary to T-wave oversensing in a biventricular ICD

Pacing Clin Electrophysiol. 2009 Jan;32(1):134-6. doi: 10.1111/j.1540-8159.2009.02187.x.

Abstract

T-wave oversensing represents a common cause of inappropriate shocks in patients with implanted cardiac defibrillators. This case report demonstrates a strategy of device programming using V-V pace delay (sequential rather than simultaneous biventricular pacing) to eliminate T-wave oversensing without decreasing sensitivity to detect true tachyarrhythmia.