To study the natural history of the hemodynamic performance of bioprosthetic heart valves, Doppler echocardiograms were recorded in a group of clinically stable patients at 2 and 5 years after replacement of native aortic valves with bioprosthetic valves. Eighteen patients completed a 2-year and 26 patients a 5-year follow-up examination. The effective orifice areas of identical models of bioprosthetic valves (Hancock II) were determined in vitro in a left-sided heart pulse duplicator system. In vivo Doppler-derived effective orifice areas were compared with the in vitro measurements for the same valve size. At both the 2- and 5-year follow-up examinations, the Doppler-derived effective orifice area was significantly less than the in vitro area (p less than 0.0001 at each interval). Ten of 16 valves evaluated serially decreased greater than 0.20 cm2 in the Doppler-derived effective orifice area between studies. The mean decrease in effective orifice area in valves evaluated serially was 0.25 +/- 0.29 cm2 (p less than 0.005). The peak transaortic gradient increased from 21 +/- 6 to 27 +/- 8 mm Hg (p less than 0.01). The mean transaortic gradient increased from 12 +/- 4 to 15 +/- 7 mm Hg (p less than 0.05). It is concluded that serial Doppler echocardiographic studies demonstrate a deterioration in the hemodynamic performance of bioprosthetic valves over time in patients with no symptoms or signs of valvular dysfunction and that Doppler echocardiography may be useful for identifying subclinical bioprosthetic valvular dysfunction.