Purpose: To analyze tangential wedged beam and field-in-field (FIF) technique doses using dose-volume histograms and conformality indices for target volume and healthy tissues within the irradiated volume.
Materials and methods: Thirty patients treated with breast-conserving surgery and postoperative whole breast radiotherapy were enrolled. Three plans were generated: a standard tangential plan with either one outer field wedge or bilateral wedges, and an FIF plan. Three indices were used: the dose homogeneity index (DHI), PTV dose improvement index (PDI(index)) and geometric conformity index (g). Also ipsilateral lung, heart and contralateral breast doses were compared for each plan.
Results: Dose homogeneity index was significantly lower for the FIF (0.117 ± 0.021) than for the single wedge (0.131 ± 0.025, p = 0.02) and double wedged plan (0.128 ± 0.025, p = 0.04), respectively. The g was significantly less in the FIF (0.70 ± 0.14) compared to the wedge plans (0.80 ± 0.17, p = 0.02 and 0.83 ± 0.16, p = 0.003). Contralateral breast doses were significantly lower in the FIF plan. The FIF plan significantly lowered MU compared to both the single wedge and bilateral wedge plans.
Conclusion: The dose distribution within the target was more homogenous, and the doses for healthy tissue were less in the FIF plan compared to the tangential wedge plans.