[Simultaneous determination of six antibiotics in disinfection products by high performance liquid chromatography tandem mass spectrometry]

Wei Sheng Yan Jiu. 2012 Mar;41(2):282-6.
[Article in Chinese]

Abstract

Objective: To develop a method for determination of the Metronidazole, Chlortetracycline hydrochloride, Oxytetracycline dihydrate, Minocycline hydrochloride, Erythromycin and Tetracycline hydrochloride in disinfection products by high performance liquid chromatography tandem mass spectrometry(LC-MS/MS).

Methods: Samples were extracted by methanol and filtered through 0.45 microm PTFE membrane filter, then analyzed by LC-MS/MS using Waters Symmetry C18 (2.1 mm x 150 mm, 3.5 microm) column in positive ion scan mode. The mobile phase was 5 mmol/L ammonium acetate, methanol and acetonitrile.

Results: The linear range was 0-2000 ng/ml and the correlation coefficients were more than 0.998, the average recoveries ranged from 74.7% to 114% with the relative standard deviations between 1.6%-20.2%. The method was successfully used to detect the content of antibiotics in 115 disinfection products.

Conclusion: The method is simple, sensitive, selective and suitable for the analysis of residual content of antibiotics in cream formulations of disinfection products.

Publication types

  • English Abstract

MeSH terms

  • Anti-Bacterial Agents / analysis*
  • Chlortetracycline / analysis*
  • Chromatography, High Pressure Liquid / methods
  • Disinfectants / analysis*
  • Metronidazole / analysis*
  • Minocycline / analysis
  • Ointments / analysis*
  • Oxytetracycline / analysis
  • Tandem Mass Spectrometry / methods

Substances

  • Anti-Bacterial Agents
  • Disinfectants
  • Ointments
  • Metronidazole
  • Minocycline
  • Chlortetracycline
  • Oxytetracycline