Reduced acoustic noise in diffusion tensor imaging on a compact MRI system

Magn Reson Med. 2018 Jun;79(6):2902-2911. doi: 10.1002/mrm.26949. Epub 2017 Oct 2.

Abstract

Purpose: To investigate the feasibility of substantially reducing acoustic noise while performing diffusion tensor imaging (DTI) on a compact 3T (C3T) MRI scanner equipped with a 42-cm inner-diameter asymmetric gradient.

Methods: A-weighted acoustic measurements were made using 10 mT/m-amplitude sinusoidal waveforms, corresponding to echo-planar imaging (EPI) echo spacing of 0.25 to 5.0 ms, on a conventional, whole-body 3T MRI and on the C3T. Acoustic measurements of DTI with trapezoidal EPI waveforms were then made at peak gradient performance on the C3T (80 mT/m amplitude, 700 T/m/s slew rate) and at derated performance (33 mT/m, 10 to 50 T/m/s) for acoustic noise reduction. DTI was acquired in two different phantoms and in seven human subjects, with and without gradient-derating corresponding to multi- and single-shot acquisitions, respectively.

Results: Sinusoidal waveforms on the C3T were quieter by 8.5 to 15.6 A-weighted decibels (dBA) on average as compared to the whole-body MRI. The derated multishot DTI acquisition noise level was only 8.7 dBA (at 13 T/m/s slew rate) above ambient, and was quieter than non-derated, single-shot DTI by 22.3 dBA; however, the scan time was almost quadrupled. Although derating resulted in negligible diffusivity differences in the phantoms, small biases in diffusivity measurements were observed in human subjects (apparent diffusion coefficient = +9.3 ± 8.8%, fractional anisotropy = +3.2 ± 11.2%, radial diffusivity = +9.4 ± 16.8%, parallel diffusivity = +10.3 ± 8.4%).

Conclusion: The feasibility of achieving reduced acoustic noise levels with whole-brain DTI on the C3T MRI was demonstrated. Magn Reson Med 79:2902-2911, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Keywords: compact 3T; diffusion tensor imaging; head-only gradient; silent MRI.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Acoustics
  • Adult
  • Anisotropy
  • Brain / diagnostic imaging
  • Diffusion Magnetic Resonance Imaging*
  • Diffusion Tensor Imaging*
  • Echo-Planar Imaging
  • Feasibility Studies
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Noise
  • Phantoms, Imaging
  • Reproducibility of Results
  • Signal-To-Noise Ratio
  • Temperature
  • Young Adult