The Accuracy of Direct Targeting Using Fusion of MR and CT Imaging for Deep Brain Stimulation of the Subthalamic Nucleus in Patients with Parkinson's Disease

J Neurol Surg A Cent Eur Neurosurg. 2021 Nov;82(6):518-525. doi: 10.1055/s-0040-1715826. Epub 2021 Feb 22.

Abstract

Introduction: In 33 consecutive patients with Parkinson's disease (PD) undergoing awake deep brain stimulation (DBS) without microelectrode recording (MER), we assessed and validated the precision and accuracy of direct targeting of the subthalamic nucleus (STN) using preoperative magnetic resonance imaging (MRI) and stereotactic computed tomography (CT) image fusion combined with immediate postoperative stereotactic CT and postoperative MRI, and we report on the side effects and clinical results up to 6 months' follow-up.

Materials and methods: Preoperative nonstereotactic MRI and stereotactic CT images were merged and used for planning the trajectory and final lead position. Immediate postoperative stereotactic CT and postoperative nonstereotactic MRI provided the validation of the final electrode position. Changes in the Unified Parkinson's Disease Rating Scale III (UPDRS III) scores and the levodopa equivalent daily doses (LEDD) and appearance of adverse side effects were assessed.

Results: The mean Euclidian distance (ED) error between the planned position and the final position of the lead in the left STN was 1.69 ± 0.82 mm and that in the right STN was 2.12 ± 1.00. The individual differences between planned and final position in each of the three coordinates were less than 2 mm. The UPDRS III scores improved by 75% and LEDD decreased by 45%. Few patients experienced complications, such as postoperative infection (n = 1), or unwanted side effects, such as emotional instability (n = 1).

Conclusion: Our results confirm that direct targeting of an STN on stereotactic CT merged with MRI could be a valid method for placement the DBS electrode. The magnitude of our targeting error is comparable with the reported errors when using MER and other direct targeting approaches.

MeSH terms

  • Deep Brain Stimulation*
  • Electrodes, Implanted
  • Humans
  • Magnetic Resonance Imaging
  • Parkinson Disease* / diagnostic imaging
  • Parkinson Disease* / therapy
  • Subthalamic Nucleus* / diagnostic imaging
  • Subthalamic Nucleus* / surgery
  • Tomography, X-Ray Computed
  • Treatment Outcome