Stereotactic depth electrode placement surgery in paediatric and adult patients with the Neuromate robotic device: Accuracy, complications and epileptological results

Seizure. 2021 Apr:87:81-87. doi: 10.1016/j.seizure.2021.03.004. Epub 2021 Mar 8.

Abstract

Objective: The number of patients requiring depth electrode implantation for invasive video EEG diagnostics increases in most epilepsy centres. Here we report on our institutional experience with frameless robot-assisted stereotactic placement of intracerebral depth electrodes using the Neuromate® stereotactic robot-system.

Methods: We identified all patients who had undergone robot-assisted stereotactic placement of intracerebral depth electrodes for invasive extra-operative epilepsy monitoring between September 2013 and March 2020. We studied technical (placement) and diagnostic accuracy of the robot-assisted procedure, associated surgical complications and procedural time requirements.

Results: We evaluated a total of 464 depth electrodes implanted in 74 patients (mean 6 per patient, range 1-12). There were 27 children and 47 adults (age range: 3.6-64.6 yrs.). The mean entry and target point errors were 1.82±1.15 and 1.98±1.05 mm. Target and entry point errors were significantly higher in paediatric vs. adult patients and for electrodes targeting the temporo-mesial region. There were no clinically relevant haemorrhages and no infectious complications. Mean time for the placement of one electrode was 37±14 min and surgery time per electrode decreased with the number of electrodes placed. 55 patients (74.3%) underwent definitive surgical treatment. 36/51 (70.1%) patients followed for >12 months or until seizure recurrence became seizure-free (ILAE I).

Conclusion: Frameless robot-guided stereotactic placement of depth electrodes with the Neuromate® stereotactic robot-system is safe and feasible even in very young children, with good in vivo accuracy and high diagnostic precision. The surgical workflow is time-efficient and further improves with increasing numbers of implanted electrodes.

Keywords: Depth electrode; Epilepsy surgery; Invasive diagnostic; Robot; Stereotactic implantation.

MeSH terms

  • Adolescent
  • Adult
  • Child
  • Child, Preschool
  • Electrodes, Implanted
  • Electroencephalography
  • Humans
  • Imaging, Three-Dimensional
  • Middle Aged
  • Robotic Surgical Procedures* / adverse effects
  • Stereotaxic Techniques
  • Young Adult