Intraoperative primary partial occlusion of the PreserFlo MicroShunt to prevent initial postoperative hypotony

Int Ophthalmol. 2023 Aug;43(8):2643-2651. doi: 10.1007/s10792-023-02664-8. Epub 2023 Mar 11.

Abstract

Purpose: The aim of the underlying study was to present a new surgical method in PreserFlo MicroShunt surgery for glaucoma. A removable polyamide suture was placed into the lumen of the MicroShunt during implantation to prevent early postoperative hypotony.

Methods: Thirty-one patients undergoing stand-alone glaucoma surgery with implantation of a PreserFlo MicroShunt and an intraluminal occlusion were retrospectively reviewed and compared to a control group without occlusion. Inclusion criteria were diagnosis of primary open-angle glaucoma or secondary open-angle glaucoma due to pseudoexfoliation or pigment dispersion. Patients with a history of filtrating glaucoma surgery were excluded.

Results: IOP decreased from 26.9 ± 6.6 to 18.0 ± 9.5 mmHg at the first postoperative day after PreserFlo MicroShunt implantation. Postoperative removal of the occluding suture resulted in a mean IOP reduction in 11.1 ± 7.6 mmHg. Mean visual acuity was 0.43 ± 0.24 logMAR during the first postoperative examination. The interval with the occluding intraluminal suture in place varied from days to 2-3 weeks. Patients were followed up to 1 year.

Conclusion: Implantation of a PreserFlo MicroShunt combined with an intraluminal suture prevented postoperative hypotony in all patients. Mean postoperative pressure was reduced despite the occluding suture in place.

Keywords: Filtering stent; Innfocus; Occlusion; PreserFlo MicroShunt; Trabeculectomy.

MeSH terms

  • Glaucoma Drainage Implants* / adverse effects
  • Glaucoma* / surgery
  • Glaucoma, Open-Angle* / etiology
  • Glaucoma, Open-Angle* / surgery
  • Humans
  • Intraocular Pressure
  • Retrospective Studies
  • Tonometry, Ocular