A Multichannel Portable Platform With Embedded Thermal Management for Miniaturized Dielectric Blood Coagulometry

IEEE Trans Biomed Circuits Syst. 2023 Aug;17(4):843-856. doi: 10.1109/TBCAS.2023.3291875. Epub 2023 Oct 6.

Abstract

This article presents a standalone, multichannel, miniaturized impedance analyzer (MIA) system for dielectric blood coagulometry measurements with a microfluidic sensor termed ClotChip. The system incorporates a front-end interface board for 4-channel impedance measurements at an excitation frequency of 1 MHz, an integrated resistive heater formed by a pair of printed-circuit board (PCB) traces to keep the blood sample near a physiologic temperature of 37 °C, a software-defined instrument module for signal generation and data acquisition, and a Raspberry Pi-based embedded computer with 7-inch touchscreen display for signal processing and user interface. When measuring fixed test impedances across all four channels, the MIA system exhibits an excellent agreement with a benchtop impedance analyzer, with rms errors of ≤0.30% over a capacitance range of 47-330 pF and ≤0.35% over a conductance range of 2.13-10 mS. Using in vitro-modified human whole blood samples, the two ClotChip output parameters, namely, the time to reach a permittivity peak (Tpeak) and maximum change in permittivity after the peak (Δϵr,max) are assessed by the MIA system and benchmarked against the corresponding parameters of a rotational thromboelastometry (ROTEM) assay. Tpeak exhibits a very strong positive correlation (r = 0.98, p < 10-6, n = 20) with the ROTEM clotting time (CT) parameter, while Δϵr,max exhibits a very strong positive correlation (r = 0.92, p < 10-6, n = 20) with the ROTEM maximum clot firmness (MCF) parameter. This work shows the potential of the MIA system as a standalone, multichannel, portable platform for comprehensive assessment of hemostasis at the point-of-care/point-of-injury (POC/POI).

MeSH terms

  • Blood Coagulation Tests
  • Humans
  • Microfluidics
  • Point-of-Care Systems*
  • Thrombelastography*