A microcomputer system for monitoring ventilation in neonates

Int J Clin Monit Comput. 1986;3(1):11-6. doi: 10.1007/BF02916393.

Abstract

The characteristics of a microcomputer-based system for measuring ventilatory parameters in neonates on mechanical ventilatory assistance are described. The tidal flows of the infant are sensed by a pneumotachometer housed in the wall of a constant pressure plethysmograph. The processed signal from the pneumotachometer is fed to a dedicated analog-to-digital converter and 6502 microcomputer, which, in turn, loads a dual ported RAM buffer with smoothed, digitized, tidal flow data. At the end of each minute the Apple 2+ processes the accumulated data and computes tidal volume, respiratory frequency, minute volume and the difference between inspired and expired volume. Numerical outputs for these variables and a histogram, representing the distribution of tidal volumes for the minute, are printed on a dot matrix printer. This inexpensive system is capable of presenting contemporary, summarized, ventilatory data in a form that is potentially very useful for the clinicians who are managing the care of the infant.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Computers*
  • Humans
  • Infant, Newborn
  • Intensive Care Units, Neonatal
  • Microcomputers*
  • Monitoring, Physiologic / instrumentation*
  • Respiration*
  • Respiration, Artificial
  • Software
  • Tidal Volume