An integrated microfluidics-tandem mass spectrometry system for automated protein analysis

Anal Chem. 1998 Sep 15;70(18):3728-34. doi: 10.1021/ac980320p.

Abstract

We describe an integrated analytical system consisting of a microfluidics device micromachined using photolithography/etching technology, a panel of computer-controlled high-voltage relays, and an electrospray ionization tandem mass spectrometer. Movement of solvents and samples on the device and off the device to the mass spectrometer was achieved by directed electroosmotic pumping induced by the activation of a suitable constellation of high-voltage relays. The system was used for the sequential automated analysis of protein digests. We demonstrate low femtomole per microliter sensitivity of detection and compatibility of the system with the automated analysis of proteins separated by two-dimensional gel electrophoresis.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Amino Acid Sequence
  • Autoanalysis
  • Electrophoresis, Gel, Two-Dimensional
  • Mass Spectrometry / methods*
  • Molecular Sequence Data
  • Proteins / analysis*
  • Sensitivity and Specificity

Substances

  • Proteins